ACCURON® 7100 Cartridge Flow Meter, Permanent: Single Range
Pipe Size Range 4 “ to 24”
Open channel design
304 Stainless Steel Cartridge
Includes fully integrated Trapezoidal Flume & Level Sensor
with IP66 Backlit LCD Digital display
Accuracy: ±1-5% of actual rate Turndown Ratio: 60:1
Engineering and plant personnel may now specify an extremely reliable and cost efficient system to monitor and measure flow in a single assembly.The Accuron Cartridge Meter (pat. pend.) shifts the responsibility of installed accuracy from the field to the factory. The added labor and expense of field programming, precise sensor alignment and costly flume installation is totally eliminated.
The Series 7100 Cartridge Meter is designed to measure flow in open channels or partially filled conduits. The Cartridge Meter utilizes ultrasonic measurments techniques to determine fluid velocity and fluid depth for calculating volume of flow.
The Series 7100 Cartridge Meters are calibrated at the factory to the application parameters provided by the customer. On-site calibration of the system is usually never required.
Each flume is correctly sized and pre-programmed for the application, and precisely aligned with a level sensor prior to field installation. This field-ready system provides the confidence that every flow monitoring device is installed correctly and accurately
ULTRASONIC LEVEL SENSOR Accuron level sensors are accurate within ±0.02”
TRAPEZOIDAL FLUME Capable of measuring very low flows, the trapezoidal cross-section provides for an extremely wide measurement range while the flat bottom design effectively clears sediment and easily passes floating debris (EPA/600/R-01/043)
INFORMATION DISPLAY
The Accuron utilizes a 160 x 128 graphical LED backlit display. The main screen can be programmed to display up to eight lines of meter information such as flow, level, velocity, time, date and relay alarms.The order of display line information is programmed to user preference. The backlit display may be programmed to be ON or OFF, or to remain OFF during a selected time interval. Display contrast is fully adjustable
DATA LOGGING
The Accuron has a built-in datalogger with eight distinct channels for logging flow and totals.The storage capacity for a single channel at 5 minute intervals is 113 days. Logger data may be visually accessed on the display of the meter in pre-programmed time intervals or retrieved through a laptop or optional modem installed within the enclosure of the meter.
Daily Averages: Daily summary allows viewing of the previous eight days.
This includes times, dates, averages, minimums, maximums and totals.
Logger Graph: In addition, a bar graph may be visually displayed on the Accuron.
The graph will display the stored logger data in pre-programmed time intervals.
Data Retrieval: Logger data can be collected by using a lapto
DATA ACQUISITION
Current Output: 4-20 mA. Isolated, 800 ohms maximum.
Relay Outputs: Three SPDT relays available for alarm conditions.
Serial Outputs: RS-232/485: With Modbus protocol.
Data Logging: Eight distinct channels are available.
QTrend Graph Package